Last Updated: May 27, 2014 04:34PM PDT
Applicable Models:  885 and 886

Problem:  The display is always showing 4080 / 6.001 even when power cycling the instrument.

Solution:  
  1. Turn off the device.
  2. Open up the battery compartment and remove the battery.
  3. From the back side facing up, locate on the right side of the battery compartment a plastic switch.  Embossed on the plastic next to the switch in the back of the battery compartment labels:RMT OFF and RMT ON
  4. Make sure the switch is set to RMT OFF.
  5. Then, insert battery and power on the device again.  It should boot back to normal measurement display.
  6. If not, power off again, wait for a few seconds, then power ON.  You can also try pressing the L/C/Z/DCR and Range Hold buttons simultaneously, then power off and on again.